# Chiller

![](/files/-LnQF3-VndLyU5rP295u)

***Periodic Maintenance*** \
\
*- the air* inlet filters on the left and right side of the chiller need to be cleaned monthly or when they appear dirty. The radiator inside the chiller behind the rear air outlet exhaust fans must also be cleaned monthly. This can be accomplished by using the compressor to blow out the inside of the chiller.\
\
\- The coolant level needs to be checked weekly or before each use. There is a coolant level gauge on the back of the unit. It should be filled until the coolant level is in the GREEN zone of the gauge. Use distilled water or other approved coolants only.

There is a 10A 250V fast acting fuse inside the power socket on the back of the unit. There is a spare fuse in the socket as well.

## Useful Links

[Manual](https://drive.google.com/file/d/1Jof2Do8vmVJm1ZrZEwuOgQ0B-mArA3jt/view?usp=sharing)

[Quick Start Guide](https://drive.google.com/open?id=19k9Gs7ne4wrSUq2_x1pxwlwmF7vBXS7R)

[Chiller Setup](https://youtu.be/w0soQMapbIU)

[Coolant Recipe](https://lasergods.com/laser-water-coolants-additives/#The_Secret_Recipe)


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://wiki.rivercitylabs.space/tools/lasers/chiller.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.